SMH Purification Biological Wastewater Treatment System Pathogen Inactivation System

The wastewater treatment of laboratory and research institutions is a challenge, and our professional equipment can be safely copened. Biological waste inactivation system, which utilizes heat and maintains high temperatures to “kill” live microorganisms, bio-waste and cellular waste.
Transform the materials available in wastewater into safety, these products can safely discharge, and will not have any negative impact on the environment;

  • Product Description
  • Technical Parameters
  • Video

Purification Biological Wastewater Treatment System Pathogen Inactivation System Description:

Smh-01 series biological wastewater treatment system is mainly used in biopharmaceutical enterprises, laboratories, scientific research institutes and other units, which discharge live toxic wastewater in the production process. These wastewater often contains part of live bacteria and viruses, which is very harmful. Pathogens must be sterilized before being discharged into the next level of sewage treatment system. Our pathogen inactivation system has the advantages of thorough sterilization, high degree of automation, simple operation, good safety and high reliability.

Purification Biological Wastewater Treatment System Pathogen Inactivation System Features:

Working principle:

The biological treatment wastewater adopts the physical thermal sterilization method, directly enters the live toxic wastewater by steam, and makes the bacteria die by denaturation of bacteria or inactivity of coagulase at high temperature. The chemical bonds in DNA and RNA of viruses absorb heat under high temperature, leading to bond fracture. Which inactivates the virus.

Waste water Treatment system is composed of collection tank and inactivated tank in a variety of forms. It is designed on site according to working conditions and user needs, equipped with appropriate pipelines and intelligent control, which can realize unattended and fully automatic operation of the whole system.

The main features of waste Water Purification System are as follows:

1. The system is mainly operated by intermittent mode. When the biological wastewater enters the collection tank through the pipe, it is inactivated by the inactivation tank.

2. All inactivated tanks are designed and manufactured according to pressure vessel standards, and collection tanks are manufactured according to atmospheric pressure vessels, which can realize in-place sterilization and cleaning. Inactivated tanks are equipped with insulation layer, which can save energy and improve product safety.

3. All valves, management and water pumps can be in place for high-temperature disinfection;

4. With waste gas treatment system;

5. PLC automatic control system, equipped with remote monitoring 485 interface, the system is equipped with a master and two auxiliary three sets of operating system, to ensure the normal operation of the system, to achieve the real unattended system;

6. With temperature, liquid level, time and other parameters setting function, can real-time display temperature, liquid level, time;

7. With overtemperature alarm, overwater alarm, operation fault alarm and other functions;

8. Have operation data records. Real-time printing and other functions;

9. With automatic water, automatic steam, automatic cooling, automatic drainage, automatic exhaust, automatic timing, automatic printing, automatic constant temperature, automatic cleaning and other functions;

 

biological wastewater treatment system,pathogen inactivation system,biological treatment wastewater,waste water treatment system,waste water purification system

Leave a Reply

Your email address will not be published. Required fields are marked *

Parameter of equipment:
Design pressure: 0.4mpa
Design temperature: 150°C
Working volume of inactivated tank: 1000L, 2000L, 3000L, 5000-10000L;
Main accessories of collecting tank:
Material Name Material Specifications Material Quantity
360 clean ball 1 “ 304 1
Differential pressure liquid level sensor 304 1
Thermometer 0~150℃ (quick-loading type) L=800 304 2
Diaphragm pressure gauge 0~ 0.6mpa (quick mounted type) 304 2
Respirator Electric heat tracing 5 “with sterilizing port 304 1
Control section:
Name Specifications Material Quantity
Japanese Yokogawa recorder 6 channels / 1
Control cabinet 800x430x1850 (h) 304 1
Control part / 1
PLC special power supply UPS power / 1